Preventing Embolism from Medical Implants by Using FlowCam Imaging Particle Analysis

Preventing Embolism from Medical Implants by Using FlowCam Imaging Particle Analysis

kurzhal-et-al-thumbnailIn a recent paper from Rostock University Medical Center, authors Anja Kurzhals et al. summarize a study they performed to determine the relevant morphological characteristics of particulates shed by a commercially available coronary stent delivery system. Flow Imaging Microscopy (FIM), also called Dynamic Image Analysis (DIA), was used to image and analyze particles produced during a simulated implantation of the cardiovascular device. 

The goal of the study was to determine which particle characteristics were most relevant in order to determine the source of the particles. Particle abrasion has the potential to cause adverse effects in patients, including embolism, and therefore particle shedding must be minimized in order for new products to be approved. New ISO guidelines require additional morphological characterization of particles prior to product launch. Discovering the morphology of the particles being shed may help determine their source. 

Kurzhals et al. used a FlowCam 8100 FIM instrument to analyze the morphology of the subvisible particles in their study. The study simulated the implantation of the coronary stent and collected the resulting particles in an aqueous suspension. FlowCam detected 858 particles in the solution, ranging in size from 10 to 500 µm. 

Histogram showing particle size distribution from Kurzhals et al.

The particle images were then organized by four different parameters (estimated spherical diameter, aspect ratio, circularity, intensity). Sorting particles by translucency (using grayscale intensity measurements) was determined to be the most effective method for ascertaining the particle source. 

FlowCam particle images from Kurzhals et al

Kurzhals et al table showing amount of particles in different intensity classesThe majority of the particulates in the sample were found to be highly translucent, causing the researchers to posit that they may originate from the lubricant coating the device's guide wires and catheters.

The team at Rostock University determined that FIM is a reliable technology for characterizing particles in this application. FlowCam makes it easy to assess the relevance of a large variety of morphological characteristics and allows the user to do so in post-processing using VisualSpreadsheet software, minimizing the required sample volume and sample runs. Unlike volumetric particle analysis techniques like Light Obscuration (LO), FIM enables researchers to determine the source of particles, and even reach conclusions about their composition, by providing statistical morphological information, and also allowing the user to see the particles for themselves.

Kurzhals et al figure showing FlowCam images of highly transparent particles and contaminants

The statistical analysis of highly-transparent particles can be challenging, particularly for technologies that collect information about particle size without collecting corresponding images (LO). FlowCam image data proved to be invaluable in identifying the morphology of the particles in this study and has the potential to do the same for many other life science applications.

Read the full paper

 

Kurzhals A, Brandt-Wunderlich C, Schmitz KP, Grabow N, Schmidt W. Systematic morphological characterization of sub-visible particles generated from cardiovascular devices using dynamic imaging analysis. Current Directions in Biomedical Engineering. 2021;7(2): 617-620. doi:10.1515/cdbme-2021-2157

Post Topics

Related Posts

Scientist pipetting sample into rack of vials
Analyzing Subvisible Particles in Protein and Cell & Gene Therapies
All types of biotherapeutics ranging from protein therapies to cell and gene therapies contain particles. While the types of particles can vary …
Read Post
Stock photo of e. coli bacteria under microscope
Identifying Bacterial Contamination in FlowCam Samples
One of the main values of using flow imaging microscopy (FIM) when analyzing pharmaceuticals is the method's sensitivity to transparent particles. …
Read Post
Need Supplies?

Find supplies and spare parts for your FlowCam instrument or ask for a quote. 

Order Now

Need Help?

Get technical support and application help. Request training or preventative maintenance.


Submit a Support Ticket

Need Information?

Check out our knowledge base including white papers, application notes, technical notes, and videos.

View Resources